|
Subject:
|
Approved by: |
Date:
|
|
Analytical X-Ray Producing Equipment Safety Requirements
|
Martin C. Graham, Ph.D.
Radiation Safety Manager |
10/2003 |
Table of Contents
- Purpose
- Equipment Requirements
- Area Requirements
- Operating Requirements
- Personnel Requirements
- Emergency Procedures
- References
-
Purpose
The following requirements are applicable to equipment utilized for
x-ray diffraction or fluorescence analysis. Analytical x-ray producing
equipment in the possession of the University of Connecticut is subject
to registration with the State of Connecticut and applicable state
and federal regulations. The University of Connecticut Radiation Safety
Office must be notified of the possession of such equipment prior
to utilization. Individuals in charge of the installation and all
equipment operators shall be familiar with applicable operating procedures
and regulations governing the x-ray installation. The Radiation Safety
Officer is empowered by the Radiation Safety Committee to immediately
terminate the operation of analytical x-ray equipment found to be
a threat to health, safety, or property until the violation is corrected.
Top of Page
-
Equipment Requirements
Requirements For An Enclosed Beam X-Ray System:
- The radiation source, beam paths, sample, detector and/or other
devices e.g. analyzing crystal, filters, etc. shall be enclosed in
a chamber, coupled chambers, beam pipes, whole system enclosure, etc.
that cannot be entered by any part of the body during normal operation.
- The inherent shielding of the chamber/enclosure walls shall be
sufficient to limit the dose rate in all regions 5 centimeters (cm)
from its outer surface to 0.25 mrem/hr during normal operations.
- The system enclosure, sample chamber, etc. closure shall be interlocked
with the x-ray tube high voltage supply and/or a shutter in the primary
beam so that no x-ray beam can enter the sample chamber while it is
open unless the interlock has been consciously and deliberately defeated.
The interlock shall be of fail-safe design.
A fail-safe characteristic means a design feature that causes beam
port shutters to close upon failure of a safety or warning device.
- If there is more than one port in the radiation source housing or
more than one radiation source, all requirements must be satisfied
for each port in every source housing associated with the system.
Safety Device: A safety device is that which prevents entry of
any portion of the individual's body into the primary x-ray beam path
or which causes the beam to be shut off upon entry into its path shall
be provided on all open-beam configurations.
An open beam system is defined as an analytical x-ray system in which
an individual could accidentally place some part of his/her body into
the primary beam path during normal operation. An analytical x-ray system
is considered to be an open-beam system unless the requirements for
an enclosed beam system outlined in Section I of this document are met.
The primary beam is the radiation that passes through an aperture
of the source housing by a direct path from the x-ray tube or a radioactive
source located in the source housing.
An analytical x-ray equipment supervisor may apply for an exemption
from the requirement of a safety device if compliance with the requirement
is not feasible. The application shall be submitted to the University
of Connecticut's Radiation Safety Officer for review and shall include
the following:
a. A description of the various safety devices that have been evaluated;
b. The reason each of these devices cannot be used; and
c. A description of the alternative methods that will be employed to
minimize the possibility of an accidental exposure, including procedures
to assure that operators and others in the area will be informed of
the absence of safety devices.
An exemption is dependant upon review and formal approval from the University
of Connecticut's Radiation Safety Officer. The x-ray equipment shall not
be utilized until approval of the exemption.
Warning Devices:
1. An open-beam configuration shall be provided with a readily discernible
indication of:
a. X-ray tube "on-off" status located near the radiation source housing,
if the primary beam is controlled in this manner; and/or
b. Shutter "open-closed" status located near each on the radiation source
housing, if the primary beam is controlled in this manner.
2. An easily visible warning light labeled with the words "X-RAY ON",
or words having similar intent shall be located near any switch that energizes
an x-ray tube and shall be illuminated only when the tube is energized.
3. Warning devices shall be labeled so that their purpose is easily identified.
The warning devices shall have fail-safe characteristics.
Ports:
Unused ports on radiation source housings shall be secured in the closed
position in a manner that will prevent casual opening.
Labeling:
All analytical x-ray equipment shall be labeled with a readily discernible
sign or signs bearing the radiation symbol and the words:
a. "CAUTION-HIGH INTENSITY X-RAY BEAM", or words having a similar intent,
on the x-ray source housing; and
b. "CAUTION RADIATION-THIS EQUIPMENT PRODUCES RADIATION WHEN ENERGIZED",
or words having a similar intent, near any switch that energizes an
x-ray tube if the radiation source is an x-ray tube; or
c. "CAUTION-RADIOACTIVE MATERIAL", or words having a similar intent,
on the source-housing if the radiation source is a radionuclide.
Top of Page
Shutters:
On open-beam configurations, each port on the radiation source housing
shall be equipped with a shutter that cannot be opened unless a collimator
or coupling has been connected to the port.
Radiation Source Housing:
Each radiation source housing shall be subject to the following requirements:
a. Each x-ray tube housing shall be equipped with an interlock that
shuts off the tube if it is removed from the radiation source housing
or if the housing is disassembled.
b. Each radioactive source housing or port cover or each x-ray tube
housing shall be so constructed that, when all shutters are closed,
the radiation measured at a distance of 5 cm from its surface is not
capable of producing a dose in excess of 2.5 mrem (0.025 mSv) in any
one hour. For systems utilizing x-ray tubes, this limit shall be met
at any specified tube rating.
Generator Cabinet:
Each x-ray generator shall be supplied with a protective cabinet that
limits leakage radiation measured at a distance of 5 cm from its surface
such that it is not capable of producing a dose in excess of 0.25 mrem
(2.5 µSv) in any one hour.
-
Area Requirements
Radiation Levels:
The local components of an analytical x-ray system shall be located
and arranged and shall include sufficient shielding or access control
such that no radiation levels exist in any area surrounding the local
component group which could result in a dose to any individual that
may be present therein in excess of the dose limits specified in applicable
University of Connecticut, state, and/or federal regulations. For
systems utilizing x-ray tubes, these levels shall be met at any specified
tube rating.
A local component is defined as part of an analytical x-ray system
and includes areas that are struck by x-rays during operation. Examples
of such components include radiation source housings, port and shutter
assemblies, collimators, sample holders, cameras, detectors, and shielding
but do not include power supplies, transformers, amplifiers, readout
devices, and control panels.
Radiation Surveys The Radiation Safety staff shall perform radiation
surveys of all analytical x-ray systems sufficient to demonstrate compliance:
- Upon installation of the equipment, and at least once
every 12 months thereafter. Annual radiation measurements shall
not be required if the Radiation Safety staff can demonstrate compliance
with applicable state and/or federal regulations;
- Following any change in the initial arrangement, number,
or type of local components in the system;
- Following any maintenance requiring the disassembly
or removal of a local component in the system;
- During the performance of maintenance or alignment
procedures if the procedures require the presence of a primary x-ray
beam when any local component in the system is disassembled or removed;
- Any time a visual inspection of the local components
in the system reveals an abnormal condition; and
- Whenever a personnel monitoring device shows a significant
increase over the previous monitoring period or the readings are
approaching the limits specified in applicable University of Connecticut,
state, and/or federal regulations.
Radiation Safety staff shall be notified immediately when the designated
laboratory supervisor of analytical x-ray equipment becomes aware of
any of the above conditions. The equipment shall not be utilized until
Radiation Safety demonstrates compliance with these requirements.
A survey meter should be kept near the equipment at all times so that
the operator may perform casual surveys frequently. It should never
be assumed that another operator or a service person left the equipment
in a safe condition.
Posting:
Each area or room containing analytical x-ray equipment shall be conspicuously
posted with a sign or signs bearing the radiation symbol and the words
"CAUTION - X-RAY EQUIPMENT" or words having a similar intent.
Top of Page
-
Operating Requirements
Procedures:
Normal operating procedures shall be written by the laboratory supervisor
for the analytical x-ray equipment and be made available to all analytical
x-ray equipment workers. No individual shall be permitted to operate
analytical x-ray equipment in any manner other than that specified in
the procedures unless such individual has received written approval
directly from the University of Connecticut's Radiation Safety Officer.
Normal operating procedures means step-by-step instructions necessary
to accomplish the analysis. These procedures shall include sample
insertion and manipulation, equipment alignment, routine maintenance,
and data recording procedures, which are all related to radiation
safety.
The radiation producing equipment supervisor shall ensure that a usage
log is maintained for each x-ray producing machine. An entry shall be
recorded on the usage log for each time the equipment is utilized. A
period of use is defined as a consecutive period of time when x-rays
are being generated. At a minimum, the following information shall be
recorded for each period of use:
- X-Ray Unit Make, Model, and Serial Number
- Operating Parameters (voltage and current)
The equipment supervisor shall utilize either key control or other administrative
controls in order to prevent unauthorized use of the x-ray equipment.
Additionally, when the x-ray unit is unattended, steps shall be taken
to prevent entry to the x-ray beam.
Bypassing:
No individual shall bypass a safety device or interlock unless such
individual has obtained prior approval from the University of Connecticut's
Radiation Safety Officer. Such approval, if granted, shall only be made
for a specified period of time. During the approved time that the safety
device or interlocks are bypassed, a readily discernible sign bearing
the words "SAFETY DEVICE NOT WORKING", or words having a similar intent,
shall be placed on the radiation source housing.
Repair or Modification of X-Ray Tube Systems:
Most severe injuries have occurred during non-routine operations such
as repair and alignment. Alignment procedures recommended by the manufacturer
of the x-ray system shall be utilized if available.
Except as specified in the section pertaining to bypassing, no operation
involving the removal of covers, shielding materials, or tube housings
or modifications to shutters, collimators, or beam stops shall be performed
without ascertaining that the tube is off and will remain off until
safe conditions have been restored. The main switch, rather than the
interlocks, shall be utilized for routine shutdown in preparation for
repairs.
If the entire system, including the x-ray tube, is under one contiguous
vacuum, and radiation leakage is less than 0.25 mrem/hr at 5 cm, and
a change in any part of the system will not increase the radiation level,
then the entire system shall be considered to be an enclosed beam system.
Radioactive Source Replacement, Testing, or Repair:
Radioactive source housings shall be opened for source replacement,
leak testing, or other maintenance or repair procedures only by individuals
authorized to specifically conduct such procedures under a license issued
by the U.S. Nuclear Regulatory Commission (NRC) or the applicable Agreement
State.
Top of Page
Personnel Requirements
Training:
No individual shall be permitted to operate or maintain analytical x-ray diffraction or spectrographic equipment until such individual has received an acceptable amount of training in radiation safety as approved by the University of Connecticut's Radiation Safety Officer and has completed the required laboratory-based training conducted by the laboratory supervisor. Competence as to the following must be demonstrated prior to operating analytical x-ray equipment:
- Identification of radiation hazards associated with the use of the equipment;
- Significance of the various radiation warning, safety devices and interlocks incorporated into the equipment, or the reasons they have not been installed on certain pieces of equipment and the extra precautions required in such cases;
- Proper operating procedures for the equipment;
- Recognition of the symptoms of an acute localized exposure; and
- Proper procedures for reporting an actual or suspected exposure.
Radiation Safety training and/or experience completed at another university and documented with the University of Connecticut's Radiation Safety Officer may be sufficient to fulfill a portion(s) of the initial training requirements.
Personnel Monitoring:
The University of Connecticut is committed to keeping exposures to radiation ALARA (As Low As Reasonably Achievable). This means that every reasonable effort shall be made to maintain radiation exposures as far below the dose limits as practical, taking into account the state of the technology, the economics of the improvements in relation to the benefits, and other socioeconomic considerations.
To monitor exposures and to ensure adherence to ALARA practices, whole body and extremity dosimetry devices shall be provided to and utilized by:
- Analytical x-ray equipment workers utilizing systems having an open-beam configuration and that are not equipped with a safety device; and
- Personnel maintaining analytical x-ray equipment if the maintenance procedures require the presence of a primary x-ray beam when any local component in the analytical x-ray system is disassembled or removed.
- Other individuals needing personnel radiation monitoring as determined by the operator's management and the University of Connecticut's Radiation Safety Officer based on regulatory requirements and exposure potential.
Responsibility:
The supervisor or person in charge of a controlled area shall be responsible for the development of standard operating procedures, the evaluation of needs, and adherence to policies with respect to radiation protection. They shall be responsible for the working conditions and for the instruction of all persons working in the area regardless of radiation hazards and methods of control. They shall also be responsible for carrying out all specified instructions and maintaining prescribed operating conditions.
All shields, interlocks, and other safety devices shall be inspected periodically and appropriately serviced. Defective shielding shall be promptly repaired and the inspection shall be repeated to determine the original degree of protection has been restored. If there is doubt about the adequacy of the repair, the Radiation Safety Officer shall be consulted.
Additionally, the laboratory supervisor shall ensure that adequate medical surveillance and radiation monitoring of personnel are carried out as necessary. If performed, a medical examination should pay particular attention to the eyes and to the skin of the hands and face.
Each operator of x-ray diffraction or spectrographic equipment shall be responsible for all operations associated with that equipment, including radiation safety. In particular:
- The operator shall, upon the instruction of Radiation Safety staff and/or responsible supervisor, follow the recommendations and instructions that have been developed in the interest of radiation protection.
- Each worker shall utilize the protective devices and radiation dosimetry provided. If the University of Connecticut's Radiation Safety Officer provides dosimetry to a worker, it shall be worn by the operator during x-ray equipment use and stored in an appropriate location in the laboratory when not in use. The dosimetry devices must be replaced with new badges at the frequency established by the Radiation Safety Committee.
A film badge shall be worn at the proper height. Both film badges and film rings shall be worn facing the radiation source.
- Each worker shall bring to the attention of the laboratory supervisor in charge any defect or deficiency in radiation protection devices, procedures or x-ray equipment function.
Each worker shall inform the Radiation Safety Officer of known or suspected abnormal radiation exposures to themselves or others.
- Each worker shall keep radiation exposures to themselves and others ALARA.
- Each operator shall be familiar with all radiation safety requirements for x-ray producing equipment and be familiar with the safety procedures as they apply to the machine he/she operates.
Emergency Procedures
If a local radiation exposure is suspected, equipment operations shall be immediately terminated, the laboratory supervisor shall be notified and the Radiation Safety Officer shall be notified at 486-3613. If a radiation emergency occurs after regular working hours, the Radiation Safety Officer may be contacted by contacting the University of Connecticut Campus Police at 911.
Arrangements may be made to receive a medical examination by Student Health Services at 486-2719. The examining physician must be informed that exposure to low-energy x-rays may have occurred.
Emergency contact information shall be posted in each analytical x-ray laboratory.
References
1. Hazards in the use of X-Ray Analytical Instrumentation, R. Jenkins and D.J. Haas. Reprinted from X-Ray Spectrometry, Vol. 2, No. 3 July 1973 with permission from John Wiley and Sons Limited. Original manuscript received 20 February 1973; accepted 13 March 1973.
2. Installations Using Non-Medical X-Ray and Sealed Gamma-Ray Sources, Energies up to 10 MeV, American National Standard, ANSI N43.3, 1993.
3. Radiation Safety for X-ray Diffraction and Fluorescence Analysis Equipment, American National Standard/Health Physics Society, ANSI/HPS N43.2, 2001.
4. Suggested State Regulations for Control of Radiation, Volume I (Ionizing Radiation), Conference of Radiation Control Program Directors, Inc., January 1991.
|